低(di)溫(wen)恒(heng)溫(wen)槽是實(shi)驗室、生產車(che)間中常(chang)用的溫(wen)度控制設(she)備,廣泛應用于化(hua)工(gong)、醫藥、食(shi)品、生物、科研等多個領域,主要用于為(wei)樣品反應、儀器檢(jian)測提供穩(wen)定(ding)的低(di)溫(wen)環境,其溫(wen)度控制的穩(wen)定(ding)性(xing)和一致性(xing),直(zhi)接(jie)影響(xiang)實(shi)驗結果和產品質(zhi)量。為(wei)避免長期(qi)使(shi)用后出現溫(wen)度偏差,確保設(she)備始終處于正(zheng)常(chang)工(gong)作狀(zhuang)態(tai),定(ding)期(qi)校準(zhun)成(cheng)為(wei)的環節。本文將詳細說明(ming)萊恩德智能、安嶼生物、海卓爾廠家低(di)溫(wen)恒(heng)溫(wen)槽的校準(zhun)方法,明(ming)確合理的校準(zhun)周(zhou)期(qi),為(wei)相關使(shi)用者提供可操作的參考依據。
一、萊(lai)恩德智能低溫(wen)恒溫(wen)槽校準前的準備工作
校準(zhun)(zhun)前做好充分準(zhun)(zhun)備(bei),可避免操(cao)作(zuo)失誤(wu),保障校準(zhun)(zhun)過程(cheng)順利進行,同(tong)時減少(shao)校準(zhun)(zhun)誤(wu)差。準(zhun)(zhun)備(bei)工作(zuo)主要包括儀器、環(huan)境(jing)、人員三(san)個方(fang)面,操(cao)作(zuo)簡單(dan)且(qie)無需(xu)復雜流程(cheng),貼(tie)合日常使用場(chang)景。
儀器準(zhun)備方面,需準(zhun)備一(yi)臺符合要求的(de)標(biao)(biao)(biao)準(zhun)溫(wen)度(du)計(ji)(優先選用標(biao)(biao)(biao)準(zhun)鉑電阻溫(wen)度(du)計(ji)或二(er)等標(biao)(biao)(biao)準(zhun)水銀溫(wen)度(du)計(ji),確保其(qi)經過(guo)正規校(xiao)(xiao)準(zhun)且在有效期內)、攪拌工(gong)具(適配(pei)(pei)恒溫(wen)槽(cao)內槽(cao)尺(chi)寸,用于(yu)使槽(cao)內介(jie)(jie)質(zhi)溫(wen)度(du)均勻)、校(xiao)(xiao)準(zhun)記(ji)錄表格(ge)(用于(yu)詳細記(ji)錄校(xiao)(xiao)準(zhun)數據(ju)、時(shi)間(jian)、環(huan)境(jing)條件等信(xin)息,便于(yu)后(hou)續追溯),同時(shi)檢查低溫(wen)恒溫(wen)槽(cao)本身,確保設(she)備無破損,槽(cao)內無雜物、污漬,加(jia)入適量適配(pei)(pei)的(de)導熱(re)介(jie)(jie)質(zhi)(如乙醇(chun)、硅油等,根據(ju)校(xiao)(xiao)準(zhun)溫(wen)度(du)范圍(wei)選擇,避免介(jie)(jie)質(zhi)凝(ning)固或揮發影響(xiang)校(xiao)(xiao)準(zhun))。
環(huan)境(jing)(jing)準備方(fang)面,校(xiao)準應在穩定的環(huan)境(jing)(jing)中進行,避免環(huan)境(jing)(jing)溫度劇烈波(bo)動、陽光直射(she)、灰塵過多(duo)或靠近熱源、冷源(如空(kong)調出風口(kou)、暖(nuan)氣、制冷設備),環(huan)境(jing)(jing)溫度控制在15℃-35℃,相對(dui)(dui)濕度不超(chao)過85%,確保環(huan)境(jing)(jing)條件不會對(dui)(dui)校(xiao)準結果產(chan)生明顯影響。
人員(yuan)準備方面,校準人員(yuan)需熟悉低溫恒溫槽的基本操(cao)(cao)作方法(fa),了(le)解校準的核心流程(cheng)和(he)注意事(shi)項,無需專(zhuan)業資質,但需認真細(xi)致(zhi),嚴格按照操(cao)(cao)作步驟進行,避免(mian)因操(cao)(cao)作不當導(dao)致(zhi)校準數(shu)據失真。
二(er)、萊恩(en)德智能低溫恒溫槽具體校準步驟
低溫(wen)(wen)(wen)恒溫(wen)(wen)(wen)槽的校(xiao)準核心是檢(jian)測其溫(wen)(wen)(wen)度(du)示值(zhi)誤(wu)差、溫(wen)(wen)(wen)度(du)均(jun)勻(yun)度(du)和溫(wen)(wen)(wen)度(du)波動(dong)度(du),三個(ge)指標的校(xiao)準步驟(zou)(zou)循序漸進(jin),操作便捷,可自行完成,無(wu)需專業(ye)檢(jian)測設備,具體步驟(zou)(zou)如下,每個(ge)環節都貼合(he)實際操作需求。
(一)溫度(du)示(shi)值誤差校準
溫(wen)度(du)(du)(du)示(shi)值(zhi)誤(wu)差是指低溫(wen)恒溫(wen)槽(cao)顯(xian)示(shi)的溫(wen)度(du)(du)(du)與實際溫(wen)度(du)(du)(du)的差值(zhi),是校準(zhun)的核心指標(biao),直接(jie)反映(ying)設(she)(she)備的溫(wen)度(du)(du)(du)控制效(xiao)果。首先,打開低溫(wen)恒溫(wen)槽(cao)電源,將設(she)(she)備溫(wen)度(du)(du)(du)設(she)(she)定為常用(yong)(yong)工作溫(wen)度(du)(du)(du)(若(ruo)設(she)(she)備有多個常用(yong)(yong)溫(wen)度(du)(du)(du)點(dian),需分別校準(zhun),一(yi)般選取3個關鍵溫(wen)度(du)(du)(du)點(dian),如低溫(wen)點(dian)、中間點(dian)、高溫(wen)點(dian),覆蓋設(she)(she)備的常用(yong)(yong)工作范(fan)圍),開啟攪拌(ban)功能,使(shi)槽(cao)內介質充分攪拌(ban),保證(zheng)溫(wen)度(du)(du)(du)均勻(yun)。
待(dai)設(she)備運行穩(wen)定(ding)后(通常需連(lian)續運行30分(fen)(fen)鐘(zhong)以上(shang),確保槽內(nei)(nei)溫(wen)(wen)(wen)度(du)達到設(she)定(ding)值且無明顯(xian)波(bo)動),將(jiang)標(biao)準(zhun)溫(wen)(wen)(wen)度(du)計(ji)(ji)緩(huan)慢(man)放(fang)入恒溫(wen)(wen)(wen)槽內(nei)(nei),使(shi)溫(wen)(wen)(wen)度(du)計(ji)(ji)的(de)感溫(wen)(wen)(wen)部分(fen)(fen)浸(jin)沒在導(dao)熱介質中,且不接(jie)觸槽壁(bi)、槽底(di)和(he)攪拌工具,避免接(jie)觸式測(ce)(ce)溫(wen)(wen)(wen)導(dao)致誤差。等(deng)待(dai)10-15分(fen)(fen)鐘(zhong),待(dai)標(biao)準(zhun)溫(wen)(wen)(wen)度(du)計(ji)(ji)讀數(shu)穩(wen)定(ding)后,同時記錄低溫(wen)(wen)(wen)恒溫(wen)(wen)(wen)槽的(de)顯(xian)示溫(wen)(wen)(wen)度(du)和(he)標(biao)準(zhun)溫(wen)(wen)(wen)度(du)計(ji)(ji)的(de)實際讀數(shu),重復(fu)測(ce)(ce)量(liang)3次,每次測(ce)(ce)量(liang)間隔5分(fen)(fen)鐘(zhong),取3次測(ce)(ce)量(liang)數(shu)據(ju)的(de)平(ping)均(jun)值作為(wei)最終(zhong)校準(zhun)數(shu)據(ju)。
最后計算溫度(du)示(shi)值(zhi)誤(wu)(wu)差,公式(shi)為:示(shi)值(zhi)誤(wu)(wu)差=恒(heng)溫槽(cao)顯示(shi)溫度(du)平均(jun)值(zhi)-標準溫度(du)計讀數平均(jun)值(zhi)。若示(shi)值(zhi)誤(wu)(wu)差在(zai)設(she)備(bei)說明書規定的允許范(fan)圍內(通常為±0.5℃-±1.0℃),則設(she)備(bei)溫度(du)示(shi)值(zhi)合格;若超出(chu)允許范(fan)圍,需對設(she)備(bei)進行調(diao)整(zheng),調(diao)整(zheng)后重新校(xiao)準,直至符合要求。
(二)溫度(du)均勻度(du)校準
溫度(du)(du)(du)(du)(du)均勻度(du)(du)(du)(du)(du)是(shi)指恒(heng)溫槽(cao)(cao)內(nei)不同位置(zhi)的溫度(du)(du)(du)(du)(du)差(cha)異,確保(bao)槽(cao)(cao)內(nei)各處溫度(du)(du)(du)(du)(du)一致,才能保(bao)證樣品受熱均勻。校準(zhun)步驟在溫度(du)(du)(du)(du)(du)示值誤差(cha)校準(zhun)的基礎上(shang)進行,當設備達到設定(ding)溫度(du)(du)(du)(du)(du)并穩定(ding)后,將標準(zhun)溫度(du)(du)(du)(du)(du)計分別(bie)放置(zhi)在恒(heng)溫槽(cao)(cao)內(nei)的5個關鍵(jian)位置(zhi)(通常(chang)為(wei)槽(cao)(cao)內(nei)中心1個點,四周各1個點,均需遠離槽(cao)(cao)壁(bi)和槽(cao)(cao)底,確保(bao)感(gan)溫部分浸沒)。
每個(ge)位(wei)置(zhi)(zhi)放置(zhi)(zhi)后(hou),等待5-10分鐘,待溫(wen)度(du)計(ji)讀(du)數穩定后(hou)記錄數據,依次(ci)完成5個(ge)位(wei)置(zhi)(zhi)的(de)測(ce)量,重復(fu)測(ce)量2次(ci),取每次(ci)測(ce)量的(de)5個(ge)位(wei)置(zhi)(zhi)數據的(de)平(ping)均值(zhi)(zhi)。計(ji)算(suan)溫(wen)度(du)均勻度(du),即5個(ge)位(wei)置(zhi)(zhi)的(de)溫(wen)度(du)平(ping)均值(zhi)(zhi)中,最(zui)a大值(zhi)(zhi)與最(zui)小值(zhi)(zhi)的(de)差值(zhi)(zhi),若差值(zhi)(zhi)不(bu)超過設(she)備允許(xu)范圍(wei)(通(tong)常(chang)為0.5℃-1.0℃),則均勻度(du)合格(ge),反之需檢查攪拌功能(neng)或(huo)調(diao)整(zheng)介質用量,重新校準。
(三)溫(wen)度波動度校(xiao)準(zhun)
溫(wen)(wen)(wen)度(du)(du)波(bo)動度(du)(du)是指恒溫(wen)(wen)(wen)槽在穩(wen)(wen)定(ding)(ding)運行時,顯示(shi)溫(wen)(wen)(wen)度(du)(du)的波(bo)動范(fan)圍,反(fan)映設備溫(wen)(wen)(wen)度(du)(du)控制的穩(wen)(wen)定(ding)(ding)性。校準(zhun)(zhun)需在設備達(da)到設定(ding)(ding)溫(wen)(wen)(wen)度(du)(du)并穩(wen)(wen)定(ding)(ding)后(hou)進行,將標準(zhun)(zhun)溫(wen)(wen)(wen)度(du)(du)計(ji)固(gu)定(ding)(ding)在槽內(nei)中(zhong)心(xin)位置,開啟攪(jiao)拌功能,連續記(ji)(ji)錄(lu)30分(fen)鐘內(nei)的溫(wen)(wen)(wen)度(du)(du)數據,每1分(fen)鐘記(ji)(ji)錄(lu)1次(ci)恒溫(wen)(wen)(wen)槽顯示(shi)溫(wen)(wen)(wen)度(du)(du)和標準(zhun)(zhun)溫(wen)(wen)(wen)度(du)(du)計(ji)讀數,共(gong)記(ji)(ji)錄(lu)30組數據。
分別計(ji)算恒溫(wen)槽顯示(shi)溫(wen)度(du)(du)和標準溫(wen)度(du)(du)計(ji)讀數(shu)的波(bo)動(dong)(dong)范(fan)圍,即每組(zu)數(shu)據中最(zui)a大值(zhi)(zhi)與最(zui)小值(zhi)(zhi)的差值(zhi)(zhi),若兩個波(bo)動(dong)(dong)范(fan)圍均在設備允許范(fan)圍內(通常(chang)為±0.1℃-±0.3℃),則溫(wen)度(du)(du)波(bo)動(dong)(dong)度(du)(du)合(he)格;若超出(chu)范(fan)圍,需(xu)檢查設備的溫(wen)控系統,排除故障后重(zhong)新校準。
三、安(an)嶼生(sheng)物(wu)低(di)溫恒溫槽校準周期說明
低溫恒溫槽的校(xiao)準(zhun)周期并(bing)非固定不變,需結合(he)設備的使用(yong)(yong)頻(pin)率、使用(yong)(yong)環境、用(yong)(yong)途等(deng)實際情況(kuang)合(he)理確定,核心原則(ze)是確保(bao)設備始終處于(yu)合(he)格(ge)狀態,避免因長(chang)期未校(xiao)準(zhun)導(dao)致溫度偏(pian)差,影響實驗或(huo)生產(chan)。結合(he)行業常規要求和實際應用(yong)(yong)經驗,校(xiao)準(zhun)周期可分(fen)為以(yi)下幾種情況(kuang),適配不同(tong)使用(yong)(yong)場(chang)景。
常(chang)(chang)規使(shi)用(yong)場景下(xia),若低溫(wen)恒溫(wen)槽使(shi)用(yong)頻率(lv)適(shi)中(zhong)(每(mei)天(tian)使(shi)用(yong)不超(chao)過8小時),使(shi)用(yong)環境穩定(無劇(ju)烈(lie)溫(wen)濕度波(bo)動(dong)、無粉塵污(wu)染),且用(yong)于普通實(shi)驗或常(chang)(chang)規生(sheng)產(非(fei)貿易結算(suan)、安全防護、醫療衛生(sheng)等(deng)關鍵(jian)領(ling)域),校(xiao)準周(zhou)期建(jian)議為12個(ge)月,即每(mei)年校(xiao)準1次,可滿足日(ri)常(chang)(chang)使(shi)用(yong)需求,同時兼顧經濟性和實(shi)用(yong)性。
特殊使用(yong)場(chang)景下,需縮短校準(zhun)周期。若設備(bei)使用(yong)頻(pin)率較(jiao)高(每天使用(yong)8小時(shi)(shi)以上,或(huo)連續24小時(shi)(shi)運行(xing)),或(huo)使用(yong)環境惡劣(溫(wen)(wen)濕度波動大、粉(fen)塵多、靠近(jin)熱源冷源),校準(zhun)周期建(jian)議(yi)(yi)縮短至6個(ge)月(yue),確(que)保(bao)及(ji)時(shi)(shi)發現(xian)溫(wen)(wen)度偏(pian)差并(bing)調(diao)整;若設備(bei)用(yong)于貿易結算(suan)、安全防護、醫(yi)療衛生(sheng)、科研檢測(ce)等關鍵領域,對(dui)溫(wen)(wen)度控制要求較(jiao)高,校準(zhun)周期建(jian)議(yi)(yi)縮短至3-6個(ge)月(yue),同(tong)時(shi)(shi)可增加期間核查(cha),每1-2個(ge)月(yue)進行(xing)1次(ci)簡單檢查(cha),確(que)保(bao)設備(bei)穩定運行(xing)。
此外,出現(xian)以(yi)下情況時(shi),需(xu)立即進(jin)行(xing)校(xiao)準,無需(xu)等待(dai)既定周期(qi):設(she)(she)備(bei)經過(guo)維修、更換(huan)核心部件(jian)(如溫(wen)控器(qi)、加熱器(qi)、傳(chuan)感(gan)器(qi))后(hou);設(she)(she)備(bei)長期(qi)閑(xian)置(超過(guo)3個月未使用(yong)(yong)),重新啟用(yong)(yong)前;發現(xian)設(she)(she)備(bei)溫(wen)度顯示異(yi)常(chang)、波(bo)動明顯,或(huo)實(shi)驗結果出現(xian)異(yi)常(chang),懷疑設(she)(she)備(bei)溫(wen)度偏(pian)差時(shi)。校(xiao)準后(hou)需(xu)妥善保存校(xiao)準記錄,留存至少(shao)2年,便(bian)于后(hou)續追溯(su)和檢查。
四、海卓爾校(xiao)準注意事項
校準過程中,需注(zhu)意以下幾點,避免操作失誤導致校準數(shu)據失真,同(tong)時保障設備(bei)和人員安全,操作簡(jian)單易執行,貼合日(ri)常使用習慣(guan)。
1. 校(xiao)準(zhun)(zhun)前,需(xu)確保標(biao)準(zhun)(zhun)溫度(du)(du)(du)計(ji)在有效期內,若超(chao)出(chu)有效期或(huo)損壞,需(xu)更換新的(de)標(biao)準(zhun)(zhun)溫度(du)(du)(du)計(ji),避(bi)免(mian)因標(biao)準(zhun)(zhun)器(qi)具不合格導(dao)致校(xiao)準(zhun)(zhun)誤差(cha);導(dao)熱介質(zhi)的(de)選擇需(xu)適配校(xiao)準(zhun)(zhun)溫度(du)(du)(du)范(fan)圍,避(bi)免(mian)低溫下介質(zhi)凝固,影(ying)響溫度(du)(du)(du)傳遞。
2. 校(xiao)準(zhun)(zhun)過程中,攪(jiao)拌(ban)工具需(xu)持續運行,確(que)保槽內介質溫(wen)(wen)(wen)度均勻(yun),避(bi)免(mian)局部溫(wen)(wen)(wen)度偏差影響(xiang)校(xiao)準(zhun)(zhun)結果;標準(zhun)(zhun)溫(wen)(wen)(wen)度計(ji)的感溫(wen)(wen)(wen)部分(fen)需(xu)浸沒在介質中,不得接觸槽壁、槽底(di)和攪(jiao)拌(ban)工具,接觸式測溫(wen)(wen)(wen)會導致數(shu)據失真。
3. 校(xiao)準(zhun)過程中,需保持環境(jing)穩定,避免隨意開關(guan)門窗、調整空調溫度,避免陽(yang)光(guang)直射恒(heng)溫槽(cao),確保環境(jing)條件不會對校(xiao)準(zhun)結(jie)果(guo)產生明顯影響;同時避免觸(chu)碰設(she)備(bei)的溫控(kong)按鈕,防(fang)止誤操作改(gai)變設(she)定溫度。
4. 校準(zhun)后,若設備(bei)溫度(du)(du)示(shi)值誤(wu)差、均(jun)勻度(du)(du)或波動度(du)(du)超(chao)出(chu)允許(xu)范(fan)圍,需及時聯系設備(bei)維(wei)修人員進行調整或維(wei)修,不得(de)繼續(xu)使用不合格設備(bei);維(wei)修后需重新(xin)校準(zhun),直至符合要(yao)求后方可(ke)投(tou)入(ru)使用。
五、總結
萊(lai)恩德智能低溫恒溫槽的校(xiao)準(zhun)(zhun)是保障設(she)備(bei)正(zheng)常(chang)運行(xing)、確(que)保實驗和(he)生產質量的重要環節(jie),校(xiao)準(zhun)(zhun)方法(fa)簡單易(yi)懂,無需(xu)專業設(she)備(bei)和(he)資質,安(an)嶼生物使(shi)用者可(ke)自(zi)行(xing)按照步驟(zou)完成,核心是做好(hao)校(xiao)準(zhun)(zhun)前準(zhun)(zhun)備(bei),規范執行(xing)示值誤差、均(jun)勻度、波動度的校(xiao)準(zhun)(zhun)流程。海(hai)卓爾校(xiao)準(zhun)(zhun)周(zhou)期需(xu)結合使(shi)用場(chang)景靈活確(que)定,常(chang)規場(chang)景每年1次,特(te)殊(shu)場(chang)景縮短(duan)至(zhi)3-6個月,出現(xian)異常(chang)情況(kuang)需(xu)立即(ji)校(xiao)準(zhun)(zhun)。
定(ding)期對低溫恒溫槽進行(xing)校準(zhun),既(ji)能(neng)及時(shi)發(fa)現并糾正溫度(du)偏差,延(yan)長設備使用(yong)壽命,也能(neng)避免(mian)因溫度(du)問題導致的(de)實(shi)驗失敗、產(chan)品(pin)不合格(ge)等(deng)損失,為實(shi)驗和生產(chan)工(gong)作提供穩定(ding)可靠的(de)溫度(du)保障(zhang)。遵循(xun)本文所述(shu)的(de)校準(zhun)方法和周期要求,可滿足大多數使用(yong)者的(de)實(shi)際需(xu)求,貼合行(xing)業(ye)常規(gui)操(cao)作規(gui)范。